MicroPette Mechanical Pipette

MicroPette Mechanical Pipette

All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.

All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.

Features for Micropette:

* Light weight, ergonomic, low force design

* Digital display clearly reads volume setting

* The pipettes cover volume range of 0.1μl to 10ml

* Easy to calibrate and maintain

* The lower part is autoclavable

* Design helps avoid repetitive strain injuries

* Manufactured from innovative materials

Features for Multi-channel Micropette:

* 8 and 12 channel pipettes are available for standard 96-well plate

* Dispensing head rotates ensure optimum pipetting convenience

* Individual piston/tip cone assemblies allow easy repair and maintenance

* Compound material tip cone design allows visual seal vertification

* Universal pipette tips can be used

Models: 

Single-channel Adjustable Volume Pipettes
Volume RangeIncrementTest Volume

(μl)

Maximume Permissible Systematic error

(Inaccuracy)

Maximume Permissible

Random error

(Inaccuracy)

%μl%μl
0.1-2.5μl0.05μl2.52.50%0.06252.00%0.05
1.253.00%0.03753.00%0.0375
0.2512.00%0.036.00%0.015
0.5-10μl0.1μl101.00%0.10.80%0.08
51.50%0.0751.50%0.075
12.50%0.0251.50%0.015
2-20μl0.5μl200.90%0.180.40%0.08
101.20%0.121.00%0.1
23.00%0.062.00%0.04
5-50μl0.5μl500.60%0.30.30%0.15
250.90%0.2250.60%0.15
52.00%0.12.00%0.1
10-100μl1μl1000.80%0.80.15%0.15
501.00%0.50.40%0.2
103.00%0.31.50%0.15
20-200μl1μl2000.60%1.20.15%0.3
1000.80%0.80.30%0.3
203.00%0.61.00%0.2
50-200μl1μl2000.60%1.20.15%0.3
1000.80%0.80.30%0.3
501.00%0.50.40%0.2
100-1000μl5μl10000.60%60.20%2
5000.70%3.50.25%1.25
1002.00%20.70%0.7
200-1000μl5μl10000.60%60.20%2
5000.70%3.50.25%1.25
2000.90%1.80.30%0.6
1000-5000μl50μl50000.50%250.15%7.5
25000.60%150.30%7.5
10000.70%70.30%3
2-10ml0.1ml10ml0.60%600.20%20
5ml1.20%600.30%15
2ml3.00%600.60%12
8-channel Adjustable Volume Pipettes
Volume RangeIncrementTest Volume

(μl)

Maximume Permissible Systematic error

(Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

%μl%μl
0.5-10μl0.1μl101.50%0.151.50%0.15
52.50%0.1252.50%0.125
14.00%0.044.00%0.04
5-50μl0.5μl501.00%0.50.50%0.25
251.50%0.3751.00%0.25
53.00%0.152.00%0.1
50-300μl5μl3000.70%2.10.25%0.75
1501.00%1.50.50%0.75
501.50%0.750.80%0.4
12-channel Adjustable Volume Pipettes
Volume RangeIncrementTest Volume

(μl)

Maximume Permissible Systematic error

(Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

%μl%μl
0.5-10μl0.1μl101.50%0.151.50%0.15
52.50%0.1252.50%0.125
14.00%0.044.00%0.04
5-50μl0.5μl501.00%0.50.50%0.25
251.50%0.3751.00%0.25
53.00%0.152.00%0.1
50-300μl5μl3000.70%2.10.25%0.75
1501.00%1.50.50%0.75
501.50%0.750.80%0.4
Fixed Volume Pipettes
Volume RangeIncrementTest Volume

(μl)

Maximume Permissible Systematic error

(Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

%μl%μl
5μl5μl1.3%0.0651.2%0.06
10μl10μl0.8%0.080.8%0.08
20μl20μl0.6%0.120.5%0.1
25μl25μl0.5%0.1250.3%0.075
50μl50μl0.5%0.250.3%0.15
100μl100μl0.5%0.50.3%0.3
200μl200μl0.4%0.80.2%0.4
250μl250μl0.4%1.00.2%0.5
500μl500μl0.3%1.50.2%1.0
1000μl1000μl0.3%3.00.2%2.0
2000μl2000μl0.3%6.00.15%3.0
5000μl5000μl0.3%150.15%7.5